Cleaning the Ulco Integrus PSV

crwdns2936025:05crwdne2936025:0 —

crwdns2935477:0crwdne2935477:0:

crwdns2935483:0crwdne2935483:0

Push canister(s) down then out of the cradle. Then dispose of the used soda lime (see suppliers recommended procedure for disposal of soda lime).

Wash all the absorber components. Do NOT use caustic cleaning fluids.

The Canister(s) and silicone shutters can be autoclaved.

The Absorber head can be put through a washer at 80 deg.C

Dry thoroughly before assembly, low pressure warm air should be passed through the head by attaching a hose to the expiratory port of the absorber.