Cardinal Health Vela Ventilator Exhalation Valve Body Replacement

crwdns2936025:015crwdne2936025:0 —

crwdns2935477:0crwdne2935477:0:

crwdns2935483:0crwdne2935483:0

Disconnect the green oxygen nebulizer tube

Remove the 2 Phillips pan-head screws behind the front panel at the lower left and right.

Disconnect the 1/8” silicone tube #4 from P2 on the exhalation differential transducer.

Disconnect the 1/8” silicone tube #5 from P1 on the exhalation differential transducer.

Disconnect the 1/8” silicone tube #1 from the turbine.