Philips Emerson CoughAssist CA Series Testing Procedure

crwdns2936025:011crwdne2936025:0 —

crwdns2935477:0crwdne2935477:0:

crwdns2935483:0crwdne2935483:0

Set the timing knobs to 5 seconds and use a stopwatch to check that each one lasts between 4.8 and 5.2 seconds.

Set the knobs to 1 second and verify that each phase lasts between 0.9 and 1.1 seconds.

If the timings are off, see “incorrect timings” on the troubleshooting guide.