ResMed AirSense 10 Pump Motor Replacement: crwdns2935265:00crwdnd2935265:01crwdnd2935265:03crwdne2935265:0 ResMed AirSense 10 Pump Motor Replacement: crwdns2935265:00crwdnd2935265:02crwdnd2935265:03crwdne2935265:0 ResMed AirSense 10 Pump Motor Replacement: crwdns2935265:00crwdnd2935265:03crwdnd2935265:03crwdne2935265:0
  • Remove the pump assembly from the pump casing and face the pump assembly with the mounting plate facing you.

  • With the pump assembly out of the pump, there is a mounting plate that needs to be removed to access the pump.

  • There are 5 rubber ridges connecting the mounting plate to the rubber sock encasing the pump.

  • Push the ridges through the holes in the mounting plate.

Entferne den Pumpeneinheit vom Pumpengehäuse und drehe diese mit der Montageplatte in deine Richtung.

Die Montageplatte muss entfernt werden, um Zugriff auf die Pumpe zu haben.

Es gibt fünf Gummirippen, die die Montageplatte mit der Gimmisocke verbinden, die die Pumpe umhüllt.

Drücke die Rippen durch die Löcher in der Montageplatte.