crwdns2933423:0crwdne2933423:0

Sostituzione motore della pompa ResMed AirSense 10

crwdns2936315:0crwdne2936315:0
crwdns2936321:0crwdne2936321:0
crwdns2931653:09crwdne2931653:0
ResMed AirSense 10 Pump Motor Replacement: crwdns2935265:00crwdnd2935265:01crwdnd2935265:03crwdne2935265:0 ResMed AirSense 10 Pump Motor Replacement: crwdns2935265:00crwdnd2935265:02crwdnd2935265:03crwdne2935265:0 ResMed AirSense 10 Pump Motor Replacement: crwdns2935265:00crwdnd2935265:03crwdnd2935265:03crwdne2935265:0
  • Position the ResMed AirSense 10 with the air outlet facing you.

  • Remove the 14mm Torx T10 screw that is located between the air outlet and the front screen.

  • Remove the wires from the pump casing's hook.

  • Carefully start to remove the pump casing from the Resmed AirSense 10 until the wire connector is exposed.

Posiziona il ResMed AirSense 10 in modo che lo scarico dell'aria sia rivolto verso di te.

Svita la vite Torx T10 da 14 mm che si trova tra lo scarico dell'aria e lo schermo anteriore.

Rimuovi i cavi dal gancio della copertura della pompa.

Inizia a rimuovere attentamente la copertura della pompa dal ResMed AirSense 10 finché non scopri il connettore del cavo.

crwdns2944171:0crwdnd2944171:0crwdnd2944171:0crwdnd2944171:0crwdne2944171:0