Reemplazo de motor de bomba en ResMed AirSense

crwdns2936315:0crwdne2936315:0
crwdns2936331:0crwdne2936331:0
crwdns2931653:09crwdne2931653:0
ResMed AirSense 10 Pump Motor Replacement: crwdns2935265:00crwdnd2935265:01crwdnd2935265:03crwdne2935265:0 ResMed AirSense 10 Pump Motor Replacement: crwdns2935265:00crwdnd2935265:02crwdnd2935265:03crwdne2935265:0 ResMed AirSense 10 Pump Motor Replacement: crwdns2935265:00crwdnd2935265:03crwdnd2935265:03crwdne2935265:0
  • Position the ResMed AirSense 10 with the air outlet facing you.

  • Remove the 14mm Torx T10 screw that is located between the air outlet and the front screen.

  • Remove the wires from the pump casing's hook.

  • Carefully start to remove the pump casing from the Resmed AirSense 10 until the wire connector is exposed.

Coloca el ResMed AirSense 10 con la salida de aire mirándote.

Retira el tornillo Torix T10 de 14mm que se encuentra entre la salida de aire y la pantalla frontal.

Retira los cables del gancho de la carcasa de la bomba

Con cuidado empieza a retirar la carcasa de la bomba del Resmed AirSense 10 hasta que el cable conector este expuesto.