Cleaning the Ulco Integrus PSV

crwdns2936315:0crwdne2936315:0
crwdns2936317:0crwdne2936317:0
crwdns2931653:05crwdne2931653:0
Cleaning the Ulco Integrus PSV: crwdns2935265:00crwdnd2935265:01crwdnd2935265:01crwdne2935265:0
  • Push canister(s) down then out of the cradle. Then dispose of the used soda lime (see suppliers recommended procedure for disposal of soda lime).

  • Wash all the absorber components. Do NOT use caustic cleaning fluids.

  • The Canister(s) and silicone shutters can be autoclaved.

  • The Absorber head can be put through a washer at 80 deg.C

  • Dry thoroughly before assembly, low pressure warm air should be passed through the head by attaching a hose to the expiratory port of the absorber.