Philips Emerson CoughAssist CA Series Testing Procedure

crwdns2936315:0crwdne2936315:0
crwdns2936321:0crwdne2936321:0
crwdns2931653:04crwdne2931653:0
crwdns2933707:0crwdne2933707:0
  • Reset the pressure knob counterclockwise, then set the control lever to inhale and check that the gauge reads between 2 and 5 cmH2O.

  • Move the control lever to exhale and note that the gauge reads between -2 and -5 cmH2O.

  • Repeat the inhale and exhale tests with the pressure knob fully clockwise, seeing if the gauge is between 57 and 63 cmH2O.

  • Release the control lever and note that the lever moves to the center and pressure drops immediately to 0 cmH2O.

  • If any of these pressures are wrong, see “limited pressure” on the troubleshooting guide.