Philips Emerson CoughAssist CA Series Testing Procedure

crwdns2935425:01crwdne2935425:0

crwdns2936315:0crwdne2936315:0
crwdns2936321:0crwdne2936321:0
crwdns2931653:01crwdne2931653:0
Philips Emerson CoughAssist CA Series Testing Procedure: crwdns2935265:00crwdnd2935265:01crwdnd2935265:01crwdne2935265:0
  • For CM models, the pressure knob is located on the left in place of the timing knobs pictured here.

  • Procedure can be found in the attached document.

  • Block the connector on the front of the machine using a rubber plug or similar item.

  • Check the pressure gauge. If it is not zero, see “pressure gauge reading is wrong” on the troubleshooting guide.

  • Set the inhale pressure knob fully clockwise, and the inhale flow and pressure knobs fully counterclockwise. On the CA models, set the manual/auto switch to manual, and set each timing knob to 1 second.