Hamilton Galileo Ventilator Power Supply Replacement

crwdns2935425:06crwdne2935425:0

crwdns2936315:0crwdne2936315:0
crwdns2936323:0crwdne2936323:0
crwdns2931653:06crwdne2931653:0
crwdns2933707:0crwdne2933707:0
  • Perform the internal voltage test linked here: page 7-5 of Section 7.3, Checking voltages.

  • If this test is not successful, it is likely that there is a loose connection somewhere between the internal power supply and the printed circuit boards. Check the wires to make sure they are securely fastened.

  • When the above test is successful, continue with the tests in the following sections:

  • 1) Section7, Checking hardware, voltages, and interface

  • 2) Either: Section 9, Running Original Upgrade 1 test software OR Section 10, Running Upgrade 2 test software