Respironics REMstar Plus M Series Blower Fan Replacement

crwdns2936025:05crwdne2936025:0 —

crwdns2935477:0crwdne2935477:0:

crwdns2935483:0crwdne2935483:0

Use the plastic opening tool as a lever to release the plastic motor frame from the retaining tab at the bottom of the module.

With the tab released, pivot the plastic frame up and away from the control module.

There are three total retaining tabs. The two on either side of the motor frame will release when pivoting the frame upward.

The small hole in the frame provides an excellent point of contact to aid in removal.

crwdns2944171:0crwdnd2944171:0crwdnd2944171:0crwdnd2944171:0crwdne2944171:0