Newport e360 Exhalation Flow Sensor Replacement

crwdns2942213:0crwdne2942213:0

  1. Newport e360 Exhalation Flow Sensor Replacement, Exhalation Flow Sensor: crwdns2935265:01crwdnd2935265:01crwdnd2935265:01crwdne2935265:0
    • Turn off ventilator before starting this repair.

    • Shut off or disconnect the air and oxygen gas supplies before starting this repair.

  2. Newport e360 Exhalation Flow Sensor Replacement: crwdns2935265:02crwdnd2935265:01crwdnd2935265:01crwdne2935265:0
    • To avoid damaging equipment, always use standard electrostatic discharge (ESD) precautions

    • Put on an ESD wrist strap or ESD gloves

  3. Newport e360 Exhalation Flow Sensor Replacement: crwdns2935265:03crwdnd2935265:01crwdnd2935265:01crwdne2935265:0
    • Open the Front Panel Door on the lower left front of the ventilator to expose Exhalation Valve EXH2105A and Flow Sensor FLS2101P.

  4. Newport e360 Exhalation Flow Sensor Replacement: crwdns2935265:04crwdnd2935265:01crwdnd2935265:01crwdne2935265:0
    • Remove the Exhalation Valve EXH2105A by releasing the retaining latch.

  5. Newport e360 Exhalation Flow Sensor Replacement: crwdns2935265:05crwdnd2935265:01crwdnd2935265:01crwdne2935265:0
    • Disconnect the Flow Sensor Cable CBL2123P from the plastic body of the Flow Sensor FLS2101P by pulling cable straight up. Do not twist.

  6. Newport e360 Exhalation Flow Sensor Replacement: crwdns2935265:06crwdnd2935265:01crwdnd2935265:01crwdne2935265:0
    • With a twisting motion, pull the plastic Flow Sensor FLS2101P away from the outlet of the Exhalation Valve EXH2105A.

crwdns2915888:0crwdne2915888:0

To reassemble your device, follow these instructions in reverse order.

crwdns2935229:02crwdne2935229:0

Lauren Ochoa

crwdns2935283:004/13/20crwdne2935283:0

165 crwdns2915208:0crwdne2915208:0

crwdns2935295:0crwdne2935295:0

crwdns2915084:0crwdne2915084:0

Cal Poly, Team S7-G21, Paton Spring 2020 crwdns2935289:0Cal Poly, Team S7-G21, Paton Spring 2020crwdne2935289:0

CPSU-PATON-S20S7G21

crwdns2931471:03crwdne2931471:0

crwdns2935297:03crwdne2935297:0

crwdns2947412:02crwdne2947412:0

I need this Exhalation Valve EXH2105A part, does anyone have a replacement?

Remigio Rodriguez - crwdns2934203:0crwdne2934203:0

I need this

Cable NEWPORT GR-CBL2123P Flow Meter Parts P/N GR-CBL2123P

Please confirmwd to edisonmanullang@gmail.com

Thanks

Edison Simanullang - crwdns2934203:0crwdne2934203:0

crwdns2917038:0crwdne2917038:0

crwdns2936625:0crwdne2936625:0:

crwdns2936751:024crwdne2936751:0 0

crwdns2936753:07crwdne2936753:0 1

crwdns2936753:030crwdne2936753:0 9

crwdns2942667:0crwdne2942667:0 597