Background

The VELA Ventilator, made in 2008, is a software driven and easy to use ventilator, with a full-function turbine gas delivery system and microprocessor control through a Graphical User Interface that provides support for pediatric to adult patients. The VELA may be configured as a conventional ventilator or non-invasive positive pressure ventilator (only available with VELA+ and VELA Comprehensive).

The VELA comes in three models: VELA, VELA+, and VELA Comprehensive. This guide is intended for the base model VELA, but may also be used with the VELA+ and VELA Comprehensive.

The VELA Ventilator is blue and has a touch interface. it is 13” W x 14.5” D x 12” H, and weighs ~38 lbs. To be sure that your ventilator is a VELA, make sure that there is the VELA marking on the front face of your device.

crwdns2915344:0crwdne2915344:0

Troubleshooting

If you are not sure what is wrong with your Cardinal Health Vela Ventilator, head over to the Cardinal Health Vela Ventilator Troubleshooting.

Specifications

IEC Classification

  • Medical Equipment: Class 1 type B

Oxygen Supply

High Pressure Connector-

  • Pressure Range: 40 to 85 psig
  • Temperature: 10 to 40°C
  • Humidity: Dew Point of gas should be 1.7°C below the ambient temperature (minimum)
  • Minimum Flow: 80 LPM at 20 psig
  • Inlet Fitting: CGA DISS-type body, No.1240

Low Pressure Connector-

  • Pressure Range: 0 to 0.5 psig
  • Maximum Flow: 80 LPM
  • Inlet Fitting: 1/4 inch tapered

Electrical Supply

AC Power Supply

  • (Nominal) / (Voltage Range) / (Frequency Range)
  • 100 VAC / 85 to 110 VAC / 47 to 65 Hz
  • 120 VAC / 102 to 132 VAC / 55 to 65 Hz
  • 230 VAC / 196 to 253 VAC / 47 to 65 Hz
  • 240 VAC / 204 to 264 VAC / 47 to 65 Hz

DC Power Supply

  • 48 VDC internal battery
  • Operates for approximately 6 hours under moderate load
  • Maximum charge time for full charge is 8 to 12 hours

Data Input / Output

Analog inputs

The ventilator provides up to 8 programmable channels for analog signal inputs. Each channel is scalable for specified ranges

  • (Range) / (Resolution)
  • 0 to 1 VDC / 0.25 mV
  • 0 to 5 VDC / 1.37 mV
  • 0 to 10 VDC / 2.5 mV

Analog Outputs-

The ventilator provides 4 signals at the analog output connector

  • Airway Pressure
    • Range: -60 to 140 cmH2O
    • Scale: 1 cmH2O / 25 mV
    • Accuracy: ±50mV or ±5% of reading, whichever is greater
    • Zero Offset: 1.5 VDC at 0 cmH2O
  • Flow
    • (Inspiratory/Expiratory)
    • When selected the ventilator provides a continuous analog voltage representative of inspiratory flow minus expiratory flow
    • Range: -100 to 200 LPM (Adult)
    • Scale Factor: 1 LPM / 10 mV (Adult)
    • 1 LPM / 25 mV (Pediatric)
    • 1 LPM / 50 mV (Neonate)
    • Accuracy: ±10% of reading or ±30 mV, whichever is greater
    • Zero Offset: 3.0 VDC at 0 LPM
    • (Machine)
    • When selected the ventilator provides a continuous analog representative of machine delivered flow.
    • Range: 0 to 200 LPM (Adult)
    • Scale Factor: 1 LPM / 25 mV (Adult)
    • 1 LPM / 50 mV (Pediatric)
    • 1 LPM / 100 mV (Neonate)
    • Accuracy: ±10% of reading or ±30 mV, whichever is greater
    • Zero Offset: None
  • Volume
    • Range: -1.00 L to 4.00 L (Adult)
    • Scale Factor: 1 L / 1 V (Adult)
    • 1 mL / 5 mV (Pediatric)
    • 1mL / 10 mV (Neonate)
    • Accuracy: ±10% of reading or ±30 mV, whichever is greater
    • Zero Offset: 1.000 VDC
  • Breath Phase
    • The ventilator provides a continuous analog voltage representative of breath phase (Inspiration = 5 VDC, Expiration = 0 VDC).

Digital Communication

  • The ventilator is constructed with one RS-232 port for bi-directional communication of data. NOTE: this is a non-operational port. Do not make connection.

Printer

  • The ventilator is constructed with a standard 25-pin female Centronics parallel printer port.

Remote Nurse Call

  • The ventilator has a modular jack configured to interface with external systems that are either wired for normally open or normally closed signals.

Video Output

  • The ventilator is constructed with a video output connector.

Atmospheric & Environmental Specifications

  • Temperature and Humidity
  • (Storage)
    • Temperature: -20 to 60°C
    • Humidity: 10 to 95% RH non-condensing
  • (Operating)
    • Temperature: 5 to 40°C
    • Humidity: 15 t0 95% RH non-condensing
  • Barometric Pressure
    • 760 to 545 mmHg

Physical Dimensions

  • Overall Size: 13” W x 14.5” D x 12” H
  • Weight: ≤ 38 lbs.

(VELA Service Manual (updated), 63-66)

Additional Information

Comparing the Vela to other ventilators

Manufacturer’s Device Page

Introduction to Operation and Controls

Vela Ventilator Diamond Series Operator’s Manual

VELA Ventilator Service manual

Jeff Stephens

crwdns2935283:002/21/11crwdne2935283:0

280,774 crwdns2915208:0crwdne2915208:0

crwdns2935297:00crwdne2935297:0

crwdns2935287:0crwdne2935287:0

iFixit crwdns2935289:0iFixitcrwdne2935289:0

Staff

crwdns2931471:0121crwdne2931471:0

crwdns2935297:072,931crwdne2935297:0

crwdns2936625:0crwdne2936625:0:

crwdns2936751:024crwdne2936751:0 0

crwdns2936753:07crwdne2936753:0 13

crwdns2936753:030crwdne2936753:0 61

crwdns2959996:0crwdne2959996:0 2,161