How to fix airflow alarm

Airflow alarm errors displays, kindly provide a technical guide on how to fix it please.

crwdns2934089:0crwdne2934089:0 crwdns2934093:0crwdne2934093:0

crwdns2934109:0crwdne2934109:0

crwdns2889612:0crwdne2889612:0 1

crwdns2947414:01crwdne2947414:0:

@dako70477 just to verify, your isolette is a Air-Shields Isolette C400QT C450QT?

crwdns2934271:0crwdnd2934271:0crwdne2934271:0

crwdns2934285:0crwdne2934285:0

crwdns2933313:01crwdne2933313:0

@dako70477 if your isolette is a Air-Shields Isolette C400QT C450QT then go to this sie Air-Shields Isolette C400QT C450QT and download the service manual. "The airflow alarm is activated by either a fan failure or a short-circuited Air Flow probe" A sensor located below deck in the normal air path of the fan controls this alarm. If air flow stops due to a fan failure, the temperature of the self-heated sensor rises causing the Air Flow alarm Indicator on the front panel to light and produce a pulsating audible tone.