How to fix airflow alarm
Airflow alarm errors displays, kindly provide a technical guide on how to fix it please.
crwdns2934109:0crwdne2934109:0
Airflow alarm errors displays, kindly provide a technical guide on how to fix it please.
crwdns2934109:0crwdne2934109:0
@dako70477 if your isolette is a Air-Shields Isolette C400QT C450QT then go to this sie Air-Shields Isolette C400QT C450QT and download the service manual. "The airflow alarm is activated by either a fan failure or a short-circuited Air Flow probe" A sensor located below deck in the normal air path of the fan controls this alarm. If air flow stops due to a fan failure, the temperature of the self-heated sensor rises causing the Air Flow alarm Indicator on the front panel to light and produce a pulsating audible tone.
crwdns2934105:0crwdne2934105:0
crwdns2936751:024crwdne2936751:0 1
crwdns2936753:07crwdne2936753:0 2
crwdns2936753:030crwdne2936753:0 3
crwdns2942667:0crwdne2942667:0 25
@dako70477 just to verify, your isolette is a Air-Shields Isolette C400QT C450QT?