How do I get this working after replacing internal battery?
I replaced the internal battery, but company won’t send updated software. I don’t care about recording data—just want the thing to work
crwdns2934109:0crwdne2934109:0
I replaced the internal battery, but company won’t send updated software. I don’t care about recording data—just want the thing to work
crwdns2934109:0crwdne2934109:0
Replacing the battery alone doesn’t seem to solve the Fault 33 error (which technically is an over pressure warning). I’m looking at replacing the transducer and will report on the results in the CPAP forum. It’s been reported that pushing a pin into the transducer eliminates the fault, I suspect it might also eliminate the unit from accurately monitoring set pressure. This would render it no longer safe for their intended purpose.
crwdns2934105:0crwdne2934105:0
crwdns2936751:024crwdne2936751:0 1
crwdns2936753:07crwdne2936753:0 5
crwdns2936753:030crwdne2936753:0 17
crwdns2942667:0crwdne2942667:0 182
How to replace internal battery?
crwdns2934271:0crwdnd2934271:0 DENNIS PRAMAS crwdne2934271:0
So it won't work after you replaced the battery?
Are you sure it's a software issue?
crwdns2934271:0crwdnd2934271:0 Hampter crwdne2934271:0