Reemplazo de la placa base del monitor de constantes vitales Welch Allyn Connex 6000

crwdns2931527:0crwdnd2931527:0crwdnd2931527:0crwdnd2931527:0crwdne2931527:0

crwdns2936315:0crwdne2936315:0
crwdns2936331:0crwdne2936331:0
crwdns2931653:03crwdne2931653:0
Welch Allyn Connex Vital Signs Monitor 6000 Motherboard Replacement: crwdns2935265:00crwdnd2935265:01crwdnd2935265:03crwdne2935265:0 Welch Allyn Connex Vital Signs Monitor 6000 Motherboard Replacement: crwdns2935265:00crwdnd2935265:02crwdnd2935265:03crwdne2935265:0 Welch Allyn Connex Vital Signs Monitor 6000 Motherboard Replacement: crwdns2935265:00crwdnd2935265:03crwdnd2935265:03crwdne2935265:0
  • Use the flat end of a spudger to "walk" the LCD cable connector (labeled J19 in the service manual) out of its socket.

  • Press first on one side of the connector, then the other to push it free. Be sure to press the connector itself, not the socket on the motherboard.

  • Pull the cable straight out of its socket.

Usa el extremo plano de un spudger para sacar el conector del cable LCD (etiquetado como J19 en el manual de servicio) fuera de su zócalo.

Presiona primero en un lado del conector, luego en el otro para liberarlo. Asegúrate de presionar el conector en sí, no el zócalo de la placa base.

Extrae el cable de su zócalo.