Reemplazo de la fuente de alimentación del monitor de constantes vitales Welch Allyn Connex 6000

crwdns2936315:0crwdne2936315:0
crwdns2936331:0crwdne2936331:0
crwdns2931653:04crwdne2931653:0
Welch Allyn Connex Vital Signs Monitor 6000 Power Supply Replacement: crwdns2935265:00crwdnd2935265:01crwdnd2935265:03crwdne2935265:0 Welch Allyn Connex Vital Signs Monitor 6000 Power Supply Replacement: crwdns2935265:00crwdnd2935265:02crwdnd2935265:03crwdne2935265:0 Welch Allyn Connex Vital Signs Monitor 6000 Power Supply Replacement: crwdns2935265:00crwdnd2935265:03crwdnd2935265:03crwdne2935265:0
  • Disconnect the ground connector from the ground plate.

  • During reassembly plug the ground cable from the IEC connector onto the ground plate in the position closest to the IEC connector. Ensure that the green ground wires are routed over the blue and brown wires.

Desconecta el conector de tierra de la placa de tierra.

Durante el reensamblaje, conecta el cable de tierra del conector IEC a la placa de tierra en la posición más cercana al conector IEC. Asegúrate de que los cables de tierra verdes se enruten sobre los cables azul y marrón.