Philips Respironics V60 Alarm Speaker Replacement

crwdns2936315:0crwdne2936315:0
crwdns2936321:0crwdne2936321:0
crwdns2931653:03crwdne2931653:0
Philips Respironics V60 Alarm Speaker Replacement: crwdns2935265:00crwdnd2935265:01crwdnd2935265:01crwdne2935265:0
  • step 3 remove the front panel

  • Separate UI from base (step 2)

  • Remove hose clamp and boot that attach blower to the gas outlet.

  • Remove the tubing from the barbed fittings inside the front panel.

  • Disconnect gas outlet and proximal port ground wires from the base.

  • Remove the screws that attach the front panel to the base and side walls.

  • NOTE: When reinstalling tubing, ensure that: • Longer (9-in.) tube is connected between manifold barb “M” (machine) to barbed gas outlet fitting. • Shorter (8-in.) tube is connected between manifold barb “P” (proximal) and barbed proximal pressure port.