After you finished the assembly of the case, look for the reset button on the PCB of the geiger counter, located near to the USB port.
-
Push the reset button of the geiger counter with your finger or with a plastic spudger to reset your device to the factory defaults (this step is necessary after battery replacement).
-
Reinsert the USB port cover into the bottom side of the geiger counter.
-
After you set the date and time and configure the datalogging settings, your device is fully functional.
-
Check the battery voltage on the screen of your device.
-
Because geiger counters are test equipments, after a battery replacement you should check the correct function of the device and you should also check the accuracy of the device with a known test source.