Gamma Scout Geiger Counter Battery Replacement, Soldering the battery to the PCB: crwdns2935265:00crwdnd2935265:01crwdnd2935265:03crwdne2935265:0 Gamma Scout Geiger Counter Battery Replacement, Soldering the battery to the PCB: crwdns2935265:00crwdnd2935265:02crwdnd2935265:03crwdne2935265:0 Gamma Scout Geiger Counter Battery Replacement, Soldering the battery to the PCB: crwdns2935265:00crwdnd2935265:03crwdnd2935265:03crwdne2935265:0
Soldering the battery to the PCB
  • Now that your battery is secured onto the PCB with the cable tie, turn the PCB over and start soldering the battery terminals onto the PCB.

  • After the solder joints have cooled down, cut down the excess part of the battery terminals with a wire cutter.

  • Be careful not to make a short-circuit with the wire cutter or with the terminal wire that will fall off after it will be cut down.

  • Make sure not to touch the upper part of the PCB or the geiger tube, because these parts of the device are under high voltage.

Ora che la batteria è fissata al PCB con una fascetta, capovolgi il PCB e inizia a saldare i terminali della batteria sul PCB.

Dopo che i giunti di saldatura si sono raffreddati, taglia la parte in eccesso dei terminali della batteria con un tagliafili.

Fai attenzione a non cortocircuitare con il tagliafili o con il filo terminale che cadrà dopo che sarà stato tagliato.

Assicurati di non toccare la parte superiore del PCB o il tubo geiger, perché queste parti del dispositivo sono sotto alta tensione.