Gamma Scout Geiger Counter Battery Replacement, Removing the old battery: crwdns2935265:00crwdnd2935265:01crwdnd2935265:03crwdne2935265:0 Gamma Scout Geiger Counter Battery Replacement, Removing the old battery: crwdns2935265:00crwdnd2935265:02crwdnd2935265:03crwdne2935265:0 Gamma Scout Geiger Counter Battery Replacement, Removing the old battery: crwdns2935265:00crwdnd2935265:03crwdnd2935265:03crwdne2935265:0
Removing the old battery
  • After you successfully removed the PCB from the case, the next step is to cut the cable tie that is holding the old lithium battery onto the PCB using a wire cutter.

  • Next you will have to remove the lithium battery from the PCB, by cutting its two terminals that are soldered onto the PCB, with a wire cutter.

  • As an alternative, you also can directly desolder the lithium battery, without cutting its terminals.

  • Make sure not to touch the upper part of the PCB or the geiger tube, because these parts of the device are under high voltage, until the battery is removed.

Dopo aver rimosso correttamente il PCB dalla custodia, il prossimo passo è quello di tagliare la fascetta ferma-cavo che tiene la vecchia batteria al litio al PCB usando un tronchese.

Successivamente, dovrai rimuovere la batteria al litio dal PCB, tagliando i due terminali saldati al PCB, con un tronchese.

Come alternativa, puoi direttamente dissaldare la batteria al litio, senza tagliare i suoi terminali.

Assicurati di non toccare la parte superiore del PCB o il tubo Geiger, perché queste parti del dispositivo sono ad alta tensione fino a quando la batteria non verrà rimossa.

crwdns2944171:0crwdnd2944171:0crwdnd2944171:0crwdnd2944171:0crwdne2944171:0