Gamma Scout Geiger Counter Battery Replacement, Removing the old battery: crwdns2935265:00crwdnd2935265:01crwdnd2935265:03crwdne2935265:0 Gamma Scout Geiger Counter Battery Replacement, Removing the old battery: crwdns2935265:00crwdnd2935265:02crwdnd2935265:03crwdne2935265:0 Gamma Scout Geiger Counter Battery Replacement, Removing the old battery: crwdns2935265:00crwdnd2935265:03crwdnd2935265:03crwdne2935265:0
Removing the old battery
  • After you successfully removed the PCB from the case, the next step is to cut the cable tie that is holding the old lithium battery onto the PCB using a wire cutter.

  • Next you will have to remove the lithium battery from the PCB, by cutting its two terminals that are soldered onto the PCB, with a wire cutter.

  • As an alternative, you also can directly desolder the lithium battery, without cutting its terminals.

  • Make sure not to touch the upper part of the PCB or the geiger tube, because these parts of the device are under high voltage, until the battery is removed.

Nachdem Sie die Platine erfolgreich aus dem Gehäuse entfernt haben, müssen Sie im nächsten Schritt den Kabelbinder, mit welchem die alte Lithium-Batterie auf der Platine fixiert ist, mit einem Drahtschneider abschneiden.

Als nächster Schritt müssen Sie die alte Lithium-Batterie von der Platine trennen, dazu müssen Sie die Axial-Lötpins, welche auf die Platine gelötet sind, mit einem Drahtschneider durchschneiden.

Alternativ können Sie die Lithium-Batterie auch direkt von der Platine entlöten, ohne die Axial-Lötpins im Voraus durchzuschneiden.

Bitte achten Sie darauf, den oberen Teil der Platine oder das Zählrohr nicht zu berühren, weil diese Teile des Gerätes unter Hochspannung stehen.