Dissolved Oxygen Node Assembly

crwdns2935425:020crwdne2935425:0

crwdns2936315:0crwdne2936315:0
crwdns2936321:0crwdne2936321:0
crwdns2931653:020crwdne2931653:0
crwdns2933707:0crwdne2933707:0
Troubleshooting
  • If the DO readings emit a reading of 0 mg/L then:

  • Make sure the DO circuit board is facing the correct orientation (refer to step 2)

  • Make sure that the jumper wires are connected correctly and securely. Likewise, check if the RX and TX wires are swapped.

  • Check the connection between the adapter cable and the housing of the DO sensor to see if it is tight and sealed.

  • Make sure there is adequate electrolyte solution in the cap of the DO sensor.

  • If the steps above fail, replace the DO sensor circuit board.

  • If the DO readings emit a reading of 100,000 mg/L then:

  • Replace the DO sensor circuit board.