+[* black] Reset the pressure knob counterclockwise, then set the control lever to inhale and check that the gauge reads between 2 and 5 cmH2O.
+[* black] Move the control lever to exhale and note that the gauge reads between -2 and -5 cmH2O.
+[* black] Repeat the inhale and exhale tests with the pressure knob fully clockwise, seeing if the gauge is between 57 and 63 cmH2O.
+[* black] Release the control lever and note that the lever moves to the center and pressure drops immediately to 0 cmH2O.
+[* black] If any of these pressures are wrong, see “limited pressure” on the troubleshooting guide.