[* icon_note] For CM models, the pressure knob is located on the left in place of the timing knobs pictured here. | |
+ | [* icon_note] Procedure can be found in the attached document. |
[* black] Block the connector on the front of the machine using a rubber plug or similar item. | |
[* black] Check the pressure gauge. If it is not zero, see “pressure gauge reading is wrong” on the troubleshooting guide. | |
[* black] Set the inhale pressure knob fully clockwise, and the inhale flow and pressure knobs fully counterclockwise. On the CA models, set the manual/auto switch to manual, and set each timing knob to 1 second. |