crwdns2933423:0crwdne2933423:0

crwdns2933803:01crwdne2933803:0

crwdns2933797:0Jaycen Kyle Mouillesseauxcrwdnd2933797:0crwdne2933797:0

crwdns2936043:0crwdne2936043:0 crwdns2933505:0crwdne2933505:0 Jaycen Kyle Mouillesseaux

crwdns2933769:0crwdne2933769:0
crwdns2933771:0crwdne2933771:0
crwdns2933801:0crwdne2933801:0

crwdns2933807:0crwdne2933807:0

[* icon_note] For CM models, the pressure knob is located on the left in place of the timing knobs pictured here.
+[* icon_note] Procedure can be found in the attached document.
[* black] Block the connector on the front of the machine using a rubber plug or similar item.
[* black] Check the pressure gauge. If it is not zero, see “pressure gauge reading is wrong” on the troubleshooting guide.
[* black] Set the inhale pressure knob fully clockwise, and the inhale flow and pressure knobs fully counterclockwise. On the CA models, set the manual/auto switch to manual, and set each timing knob to 1 second.