crwdns2933423:0crwdne2933423:0

crwdns2933803:09crwdne2933803:0

crwdns2933797:0Grace Giuscrwdnd2933797:0crwdne2933797:0

crwdns2936043:0crwdne2936043:0 crwdns2933505:0crwdne2933505:0 Grace Gius

crwdns2933769:0crwdne2933769:0
crwdns2933771:0crwdne2933771:0
crwdns2933801:0crwdne2933801:0

crwdns2933807:0crwdne2933807:0

[* black] Pressure accuracy test is complete.
-[* black] If Inhalation/Exhalation pressure readings are out of range at 100 cmH2O, recommended troubleshooting and repair is as follows:
-[* black] 1. Check for leaks at circuit connections, test lung, filters, etc.
-[* black] 2. Check for kinked or cut tubing from inhalation module to SOL4, and from SOL4 to sensor PCB.
-[* black] 3. Check for kinked or cut tubing from exhalation pressure tap to SOL3, and from SOL3 to sensor PCB.
-[* black] 4. Check for leaks at the oxygen sensor/oxygen valve connection.
-[* black] 5. Replace the sensor PCB.
-[* black] 6. Replace the 3-station solenoid.