+[* black] Pressure accuracy test is complete.
+[* black] If Inhalation/Exhalation pressure readings are out of range at 100 cmH2O, recommended troubleshooting and repair is as follows:
+[* black] 1. Check for leaks at circuit connections, test lung, filters, etc.
+[* black] 2. Check for kinked or cut tubing from inhalation module to SOL4, and from SOL4 to sensor PCB.
+[* black] 3. Check for kinked or cut tubing from exhalation pressure tap to SOL3, and from SOL3 to sensor PCB.
+[* black] 4. Check for leaks at the oxygen sensor/oxygen valve connection.
+[* black] 5. Replace the sensor PCB.
+[* black] 6. Replace the 3-station solenoid.