+ | [title] charging |
---|---|
+ | [* black] Connect the ventilator back to the AC power. |
+ | [* red] Make sure the battery is properly installed by verifying that the yellow Battery (charged) LED on the front panel flashes. The flashing LED indicates the battery is being charged. |
+ | [* icon_note] A new battery must be charged for at least 5 hours before being placed into service. |
+ | [* icon_caution] Following battery installation, if a ***Check Vent*** or ***Vent Inoperative*** alarm occurs during the preoperational check, discontinue use of the ventilator immediately and contact Respironics. The ***Vent Inoperative*** alarm occurs if AC power is disconnected and a battery is not installed, or if the battery is fully discharged. |