+ | [title] charging |
---|---|
+ | [* black] Connect the ventilator back to the AC power. |
+ | [* red] Make sure the battery is properly installed by verifying that the yellow ***Battery (charged) LED*** on the front panel flashes. The flashing LED indicates the battery is being charged. |
+ | [* icon_note] A new battery must be charged for at least 5 hours before being placed into service. |
+ | [* icon_note] The manufacturer recommends that the ventilator’s batteries be fully charged before ventilating a patient. If the batteries are not fully charged and AC power fails, always pay close attention to the level of battery charge. |
+ | [* icon_note] The ventilator charges the battery whenever the ventilator is connected to AC, with or without the ventilator switched on. The ***Battery (charged) LED*** flashes to show that the battery is being charged. |