crwdns2933803:020crwdne2933803:0
crwdns2933797:0Jonathan Niecrwdnd2933797:0crwdne2933797:0
crwdns2936043:0crwdne2936043:0 crwdns2933505:0crwdne2933505:0 shingle
- crwdns2933769:0crwdne2933769:0
- crwdns2933771:0crwdne2933771:0
- crwdns2933801:0crwdne2933801:0
crwdns2933807:0crwdne2933807:0
+ | [title] Troubleshooting |
---|---|
+ | [* icon_caution] If the DO readings emit a reading of 0 mg/L then: |
+ | [* black] Make sure the DO circuit board is facing the correct orientation (refer to step 2) |
+ | [* black] Make sure that the jumper wires are connected correctly and securely. Likewise, check if the RX and TX wires are swapped. |
+ | [* black] Check the connection between the adapter cable and the housing of the DO sensor to see if it is tight and sealed. |
+ | [* black] Make sure there is adequate electrolyte solution in the cap of the DO sensor. |
+ | [* icon_note] If the steps above fail, replace the DO sensor circuit board. |
+ | [* icon_caution] If the DO readings emit a reading of 100,000 mg/L then: |
+ | [* black] Replace the DO sensor circuit board. |