Introduction
This guide is created for the replacement of oxygen cells in the Hamilton-T1 Ventilator. Replacing the oxygen cells is a crucial step in maintaining and keeping a working ventilator.
'''NOTE:'''
* Replace the oxygen cell with genuine Hamilton Medical parts only; otherwise, oxygen measurement will not function. To prevent a permanent alarm make sure to use special Hamilton Medical oxygen cells only.
* To prevent leakage within the ventilator, make sure an oxygen cell is installed at all times, even if you use an external monitor or disable oxygen monitoring.
* To prevent a permanent alarm use special Hamilton Medical oxygen cells only.