Conclusion |
If Inhalation/Exhalation pressure readings are out of range at 100 cmH2O, recommended troubleshooting and repair is as follows and can be found in the [https://www.ifixit.com/Document/uHU5MIrC51dBUcNN/Philips+Respironics+V200+Espri.pdf|Service Manual] in Chapter 8.7.5 page 8-44: | | # Check for leaks at circuit connections, test lung, filters, etc. | # Check for kinked or cut tubing from inhalation module to SOL4, and from SOL4 to sensor PCB. | # Check for kinked or cut tubing from exhalation pressure tap to SOL3, and from SOL3 to sensor PCB. | # Check for leaks at the oxygen sensor/oxygen valve connection. | # Replace the sensor PCB. | # Replace the 3-station solenoid. |
|