crwdns2915892:0crwdne2915892:0
In order to replace most other parts in the Newport e500, the removal of the right side panel is required. In order to perform this operation, a Philips screwdriver and torque wrench will be required. For more information refer to page 41 of the service manual.
crwdns2942213:0crwdne2942213:0
-
-
Place device so that the top cover is facing you.
-
Open the top cover.
-
-
-
Remove the four hex nuts and lock washers (two at each hinge) that secure the cover to the hinges on the left side of the pneumatic box.
-
-
-
Remove the cover and set aside. Retain hardware for re-assembly.
-
-
-
Carefully place the GDU on its left side.
-
-
-
Remove the four screws and washers at the top edge of the right side panel.
-
-
-
Remove the two screws and washers from the underside of the right side panel.
-
-
-
Remove the right side panel. Retain hardware for re-assembly.
-
To reassemble your device, follow these instructions in reverse order.
To reassemble your device, follow these instructions in reverse order.