If you need to access important parts associated with the Newort e500, removal of the exhalation module is an integral step. This step serves as a common pre-requisite to completing other guides. Make sure you are careful with the outlet boot as it is fragile. For more information refer to page 42 of the service manual.
-
-
Pull up and remove the flexible rubber exhalation outlet boot (push fit).
-
-
-
Disconnect the tubes from both sides of the flow sensor and the two tubes from the exhalation valve.
-
-
-
Pull the exhalation module straight up to remove it from the GDU.
-
To reassemble your device, follow these instructions in reverse order.