Please refer to page 7 of the device’s Technical Manual for source information.
-
-
Set the gas supply to 5 L/min.
-
-
-
Completely close the Peak Inspiratory Pressure control and the Max Pressure control by turning both knobs completely clockwise.
-
-
-
Close the bleed valve and check that the gauge reads at least 60 cmh2O.
-
-
-
Close the bleed valve and check that the gauge reads no higher than 80 cmH2O.
-
-
-
Close the bleed valve and turn the Max Pressure Relief valve until the manometer reads 40 cmh2O.
-
-
-
Check that the manometer needle rises and falls smoothly.
-
-
-
Gently rotate the Max Pressure Relief cover over the Max Pressure Relief knob.
-
-
-
Reset the peak inspiratory pressure to 20 cmh2O and turn off the gas flow. Testing is now complete. If the F&P Neopuff/Perivent Infant Resuscitator fails any of these tests, the valve assembly should be regarded as faulty and replaced with a new valve assembly (see the parts list (section 4.1.2) for part numbers).
-
To reassemble your device, follow these instructions in reverse order.