Please refer to page 7 of the device’s Technical Manual for source information.
-
-
Adjust the gas flow to 10 L/min.
-
-
-
Close the Peak inspiratory Pressure valve by turning the knob fully clockwise.
-
-
-
Adjust the Max Pressure Relief knob clockwise or counterclockwise until manometer reads 40 cmH20
-
-
-
Turn the Peak Inspiratory Pressure knob counterclockwise so the manometer reads 20 cmH20 and shut off the gas flow.
-
Alternate settings for the Max Pressure Relief should be made in accordance with hospital protocol.
Alternate settings for the Max Pressure Relief should be made in accordance with hospital protocol.