Hamilton Medical Ventilator Replacing the LCD

crwdns2911024:0crwdne2911024:0
crwdns2911026:0crwdne2911026:0

crwdns2942213:0crwdne2942213:0

  1. Hamilton Medical Ventilator Replacing the LCD, LCD: crwdns2935265:01crwdnd2935265:01crwdnd2935265:01crwdne2935265:0
    • First open the main enclosure by removing the four hexagonal-drive (Allen) screws at the rear of the enclosure.

  2. Hamilton Medical Ventilator Replacing the LCD: crwdns2935265:02crwdnd2935265:01crwdnd2935265:01crwdne2935265:0
    • Unscrew the control board mounting plate.

  3. Hamilton Medical Ventilator Replacing the LCD: crwdns2935265:03crwdnd2935265:01crwdnd2935265:01crwdne2935265:0
    • “Hinge” back the mounting plate, complete with control board. This exposes the LCD.

  4. Hamilton Medical Ventilator Replacing the LCD: crwdns2935265:04crwdnd2935265:01crwdnd2935265:01crwdne2935265:0
    • Unplug the cable from the GMP adaptor board from the LCD.

  5. Hamilton Medical Ventilator Replacing the LCD: crwdns2935265:05crwdnd2935265:01crwdnd2935265:01crwdne2935265:0
    • Unplug from the dc/ac board, the cable to the interface board.

  6. Hamilton Medical Ventilator Replacing the LCD: crwdns2935265:06crwdnd2935265:01crwdnd2935265:01crwdne2935265:0
    • Unscrew the four hexagonal-drive (Allen) screws securing the two brackets on which the LCD is mounted.

  7. Hamilton Medical Ventilator Replacing the LCD: crwdns2935265:07crwdnd2935265:01crwdnd2935265:01crwdne2935265:0
    • Lift out the complete assembly.

  8. Hamilton Medical Ventilator Replacing the LCD: crwdns2935265:08crwdnd2935265:01crwdnd2935265:01crwdne2935265:0
    • Unplug both of the lights from the dc/ac board.

  9. Hamilton Medical Ventilator Replacing the LCD: crwdns2935265:09crwdnd2935265:01crwdnd2935265:01crwdne2935265:0
    • Remove the LCD from the two brackets, by removing the black hexagonal-drive (Allen) screws on the LCD. NOTE: Do not remove the silver hexagonal-drive (Allen) screws on the SIDE of the LCD.

  10. Hamilton Medical Ventilator Replacing the LCD: crwdns2935265:010crwdnd2935265:01crwdnd2935265:01crwdne2935265:0
    • When fitting a new display, there must be an offset of approximately 2 mm between the plastic sealing strips on the LCD, and the brackets.

  11. Hamilton Medical Ventilator Replacing the LCD: crwdns2935265:011crwdnd2935265:01crwdnd2935265:01crwdne2935265:0
    • If necessary, adjust this gap using the silver screws on the side of the LCD.

    • The surface of the LCD must be clean. Wipe it free of finger marks and dust with an antistatic paper towel.

  12. Hamilton Medical Ventilator Replacing the LCD: crwdns2935265:013crwdnd2935265:01crwdnd2935265:01crwdne2935265:0
    • The cable from the GMP to the LCD must be in the cable grip on the control board mounting plate.

    • When screwing the LCD assembly into GALILEO, you must align it with the glass screen in GALILEO. Check this by looking at the front of GALILEO.

    • Turn on GALILEO, and make sure that the LCD lights up, and is properly aligned.

crwdns2915888:0crwdne2915888:0

To reassemble your device, follow these instructions in reverse order.

Sofya Shatalova

crwdns2935283:004/07/20crwdne2935283:0

229 crwdns2915208:0crwdne2915208:0

crwdns2935295:0crwdne2935295:0

crwdns2915084:0crwdne2915084:0

Cal Poly, Team S7-G26, Paton Spring 2020 crwdns2935289:0Cal Poly, Team S7-G26, Paton Spring 2020crwdne2935289:0

CPSU-PATON-S20S7G26

crwdns2931471:03crwdne2931471:0

crwdns2935297:04crwdne2935297:0

crwdns2947412:00crwdne2947412:0

crwdns2917038:0crwdne2917038:0

crwdns2936625:0crwdne2936625:0:

crwdns2936751:024crwdne2936751:0 1

crwdns2936753:07crwdne2936753:0 1

crwdns2936753:030crwdne2936753:0 13

crwdns2942667:0crwdne2942667:0 308