crwdns2915892:0crwdne2915892:0
For more information on removing the safety valve block, please refer to section 10 in the service manual.
crwdns2942213:0crwdne2942213:0
-
-
Remove 4 Hex Allen screws (A) from the Rail surrounding the Ventilation Unit.
-
-
-
The Standoff Brackets (C) remain with the Main Cover.
-
-
-
Lift the Main Cover (A) from the Ventilation Unit using the Standoff Brackets (B) attached to the Main Cover.
-
-
-
Remove Safety Valve Block
-
1. Disconnect the cable from the Ambient Valve Solenoid
-
2. Disconnect the Shaped Hose (A)
-
3. Remove the 2 Hex Allen Screws (B)
-
4. Remove the Safety Valve Block
-
To remove Ambient Valve Solenoid: Unscrew the Ambient Valve Solenoid (C) form the Safety Valve Block
-
To reassemble your device, follow these instructions in reverse order.
To reassemble your device, follow these instructions in reverse order.