This guide will walk you through the process of replacing the Main Power Receptacle in a Bear 1000 Ventilator. Found on page 89 in the full service manual.
crwdns2942213:0crwdne2942213:0
-
-
Remove the ribbon cable at the connection on the oxygen blender control board.
-
-
-
Mark the plastic hoses attached to the oxygen blender (so you can put them back on the same way) and then remove.
-
-
-
The air/oxygen blender has two screws close to the control board. Remove them and remove the unit.
-
-
-
Remove the pneumatic and electrical connections labeled.
-
-
-
Remove the water trap bowls on the rear of the unit.
-
-
-
Remove the four screws and remove the manifold.
-
-
-
Remove the two screws holding the power supply to the manifold.
-
-
-
Note and mark the orientation of the wires. Remove the two wires going to the fuses at the fuse end and the ground wire going to the main ground stud.
-
-
-
Using an allen wrench, remove the mounting screws and nuts from the receptacle.
-
-
-
Transfer the three wires form the old receptacle to the new one
-
To reassemble your device, follow these instructions in reverse order.