Autosampler Programming (For Old & New Versions)

    1. Autosampler Programming  (For Old & New Versions), Turn on battery and autosampler: crwdns2935265:01crwdnd2935265:01crwdnd2935265:02crwdne2935265:0 Autosampler Programming  (For Old & New Versions), Turn on battery and autosampler: crwdns2935265:01crwdnd2935265:02crwdnd2935265:02crwdne2935265:0
      • 1. Switch battery "ON".

      • 2. Autosampler: Push "ON" button.

      • It will say ...STANDBY...

      crwdns2952109:0crwdne2952109:0

      crwdns2917038:0crwdne2917038:0

    2. Autosampler Programming  (For Old & New Versions), Select Program Mode: crwdns2935265:02crwdnd2935265:01crwdnd2935265:02crwdne2935265:0 Autosampler Programming  (For Old & New Versions), Select Program Mode: crwdns2935265:02crwdnd2935265:02crwdnd2935265:02crwdne2935265:0
      • 3. Push "ENTER/PROGRAM" button to see the "[PROGRAM, CONFIGURE] SAMPLER" screen. Push “ENTER/PROGRAM” button to select "PROGRAM".

      • 4. This will bring up the "[TIME, FLOW] PACED SAMPLING" screen. Use arrow to select "FLOW" then push “ENTER/PROGRAM” button.

      crwdns2952109:0crwdne2952109:0

      crwdns2917038:0crwdne2917038:0

    3. Autosampler Programming  (For Old & New Versions), Set Number, Type, and Size of Samples: crwdns2935265:03crwdnd2935265:01crwdnd2935265:03crwdne2935265:0 Autosampler Programming  (For Old & New Versions), Set Number, Type, and Size of Samples: crwdns2935265:03crwdnd2935265:02crwdnd2935265:03crwdne2935265:0 Autosampler Programming  (For Old & New Versions), Set Number, Type, and Size of Samples: crwdns2935265:03crwdnd2935265:03crwdnd2935265:03crwdne2935265:0
      • 5. New screen shows: "SAMPLE EVERY __ PULSES." Press 20. Then push the “ENTER/PROGRAM” button.

      • 6. New screen will read: "MULTIPLEX SAMPLES?" Use arrow to select "NO". Then push the “ENTER/PROGRAM” button.

      • 7. New screen will read: "SAMPLE VOLUMES OF "____" ML". Press 1000. Then push the “ENTER/PROGRAM” button.

      crwdns2952109:0crwdne2952109:0

      crwdns2917038:0crwdne2917038:0

    4. Autosampler Programming  (For Old & New Versions), Starting Sampling: crwdns2935265:04crwdnd2935265:01crwdnd2935265:03crwdne2935265:0 Autosampler Programming  (For Old & New Versions), Starting Sampling: crwdns2935265:04crwdnd2935265:02crwdnd2935265:03crwdne2935265:0 Autosampler Programming  (For Old & New Versions), Starting Sampling: crwdns2935265:04crwdnd2935265:03crwdnd2935265:03crwdne2935265:0
      • 8. New screen will read: "ENTER START TIME?" Use arrow to select "NO". Then push the “ENTER/PROGRAM” button. New screen: "PROGRAMMING SEQUENCE COMPLETE...". Then screen will read: "PROGRAM HALTED".

      • 9. Push "START SAMPLING" button. Screen will read: "[START, RESUME] SAMPLING PROGRAM?" Push "ENTER/PROGRAM" button to "START". Then push "ENTER/PROGRAM" button again to "START SAMPLING AT BOTTLE 1".

      • 10. Then should see on screen: "START FLOW COUNT AT __:__" (time).

      • It will take some time before it starts sampling.

      • Eventually screen will show: " BOTTLE 1 AFTER 20 PULSES".

      crwdns2952109:0crwdne2952109:0

      crwdns2917038:0crwdne2917038:0

crwdns2935217:0crwdne2935217:0

crwdns2935221:0crwdne2935221:0

crwdns2935227:0crwdne2935227:0

Brooke Mason

crwdns2935283:005/29/19crwdne2935283:0

2 958 crwdns2915208:0crwdne2915208:0

crwdns2935297:017crwdne2935297:0

crwdns2935287:0crwdne2935287:0

Open-Storm crwdns2935289:0Open-Stormcrwdne2935289:0

Community

crwdns2931471:014crwdne2931471:0

crwdns2935297:0275crwdne2935297:0

crwdns2947412:00crwdne2947412:0

crwdns2917038:0crwdne2917038:0

crwdns2936625:0crwdne2936625:0:

crwdns2936751:024crwdne2936751:0 3

crwdns2936753:07crwdne2936753:0 23

crwdns2936753:030crwdne2936753:0 57

crwdns2942667:0crwdne2942667:0 569