Background
The Sherwood FirstTemp Genius 3000A is a handheld battery operated thermostat that is used in the medical field. This device is designed to make 2 second readings without inserting within the body. Depending on where the device is used on the body, temperature measurements can be changed to either read in Fahrenheit or Celsius.
This device comes equipped with indicators to provide you with information of the temperature displayed in tenths(0.1), temperature display using calibrator in hundredths(0.01), diagnostic information, service menu, errors, battery condition, settings, and temperature mode.
Controls you'll find on the device would be the "Scan" button and "Mode" Button. When using the mode button you will see more controls such as CALIB (Calibration), SPC (Special Function), EQUIV, LOCK, and DEGREES F/ DEGREES C.
Additionally to the thermostat there is a wall mounting kit, blackbody for water baths, base assembly, 9v alkaline battery, Coiled assembly cords, calibrator, thermostat stand, as well as operation and service manual.
Identification
The Sherwood FirstTemp Genius 3000A is encased within a white and blue ABS plastic housing. The dimensions of the device are 8.5"(length) x 2.2"(width) x 3.0"(height). The base model variant has no electrical connections to the probe and its measurements are 9.2" (length) x 2.7"(width) x 2.0" (height). For the base model there is an optional security code that can be attached to the thermostat. To identify the model number, remove the handheld wand from its base and the model number will be printed in the top right corner of writing on the wand.
Technical Specifications
Calibrated Accuracy: +0.2°F (+0.1°C) 98°F-102°F. Based on ASTM standard
E1112-86.
Calibrated Temperature Range: 85°F-110°F (29.4°C-43.3°C).
Ambient Temperature Operating Range: 60°F-95°F (15.6°C-35°C).
Storage Temperature Range: -4°F to 122°F (-20°C to 50°C).
Response Time: Approximately 1-2 seconds.
Primary Modes: Tympanic and Timer.
Resolution: 0.1°F or 0.1°C displayed on LCD.
Power: One 9-volt alkaline battery.
Battery Life: Approximately 5,000 temperatures.
Two-Stage Battery Warning: At Stage 1, replace battery as soon as practical.
At Stage 2, replace battery immediately or thermometer will not operate.
Temperature Conversion: User-programmable for rectal, oral or core
equivalences (°F or °C).
Unique Features: Incorporates 15-second pulse timer. Instant In-Service Menu
System prompts on method of use via LCD screen. Calibration may be
checked at user site with Genius calibrator.
Size:
Probe = 8.5"L x 2.2"W x 3"H
Base = 9.2"L x 2.7"W x 2"H
Weight:
Probe = 7 oz.
Base = 11 oz.
"Sherwood FirstTemp Genius 3000A Service Manual page 25"